HYCOUNT 5

HYCOUNT 5

REF : HYCOUNT_5

5-Part differential analyzer – Open tube

• 5-part differentiation, 27 parameters, 3 DIFF scatter grams and 1 BASO scatter gram, 3 histograms for WBC, RBC and PLT.
• Tri-angle semi-conductor Laser scatter combined with chemical dye method, advanced flow cytometry
• Compact, powerful and affordable
• Only 15ul sampling volume
• Up to 60 samples per hour
• 3 counting modes: venous blood, capillary blood and pre-diluted
• Capability to flag abnormal samples
• Large storage capacity: up to 50,000 samples
• Reagents protected by RFID card.

HYCOUNT 5

HYCOUNT 5

5-Part differential analyzer – Open tube

• 5-part differentiation, 27 parameters, 3 DIFF scatter grams and 1 BASO scatter gram, 3 histograms for WBC, RBC and PLT.
• Tri-angle semi-conductor Laser scatter combined with chemical dye method, advanced flow cytometry
• Compact, powerful and affordable
• Only 15ul sampling volume
• Up to 60 samples per hour
• 3 counting modes: venous blood, capillary blood and pre-diluted
• Capability to flag abnormal samples
• Large storage capacity: up to 50,000 samples
• Reagents protected by RFID card.

SPECIFICATIONS

REFERENCE :

HYCOUNT_5

PRINCIPLES :

• Impedance method for WBC, RBC, PLT counting
• Cyanide-free reagent for Hemoglobin test
• Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel

THROUGHPUT :

Up to 60 samples per hour

SAMPLE VOLUME :

Whole Blood: 15ul, Pre-diluted: 20ul

TEST MODE :

CBC CBC+DIFF

DATA STORAGE CAPACITY :

Up to 50,000 results including numeric and graphical information

COMMUNICATION :

LAN port supports HL7 protocol

PERFORMANCES

PARAMETERCARRYOVERREPEATABILITYLINEARITY
WBC ≤0.5%≤2.0% (4-15×10 puissance 9 /L)0.00-300×10 puissance 9 /L
RBC ≤0.5%≤1.5% (3.5-6.0×10 puissance 12 /L)0.00-8.5×10 puissance 12 /L
HGB ≤0.5%≤1.5% (110-180 g/L)0-250 g/L
PLT ≤1.0%≤4.0% (150-500×10 puissance 9 /L)0.00-3000×10 puissance 9 /L

PARAMETERS

• WBC, Lym%, Lym#,Mon%, Mon#,Neu%, Neu#,Eos%, Eos#, Bas%, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT, and 4 research parameters LIC%, LIC#, ALY%, ALY#
• 3 DIFF scatter grams and 1 BASO scatter gram, 3 histograms histogrammes

◼︎ CONTACT US NOW ◼︎